Chapter 5. Troubleshooting
Problems
The pump stops.
Red ALARM LED lights up.
The roller pump heads do
not rotate.
There is no irrigation.
The irrigation pump starts
spinning frantically, emitting
excessive noise.
Cause
The HIGH PRESSURE safety
has activated.
The LOW PRESSURE safety
has activated: �Pres. L� blinks.
The hydrophobic filter is wet.
There is no more
irrigation fluid.
The power is on �O� or is
on Stop.
The stop-cock on the
sheath or a clamp is closed,
stopping outflow.
There is no more
irrigation fluid.
29
Possible solutions
Lower the joint pressure.
Press the Run/Stop pad to
restart the pump.
Verify that the luer-lock is
tightly closed.
Press the Fill Chamber to
stop the alarm, and press
the Run/Stop pad to restart
the pump.
Replace the �One Day Set™�.
Spike the bag and press the
Fill Chamber pad.
Check that the main power
switch on the back of the
pump is on the �I� position,
and that the Run/Stop pad
green light is on.
Check the inflow line from the
bags to the joint. Open any
closed clamps or stop-cocks.
Replace the saline solution
bags and/or check the
clamps under the bags.
Press Fill Chamber if chamber
is lower than 1/3.