Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 152

Pressure Transducer Test

Procedure:
• S witch the FMS DUO+ on "I" on the back panel.
• P ress the "POWER" pad (4).
• P ress "Run/Stop" pad (18).
• R aise both covers.
• C onnect the manometer (with pressure bulb
attached) to the Pressure transducer connection
luer-lock (19).
• P ressure reference = 50 (+/- 5 feet H20).
• M easure real pressure with the manometer and
note the value.
• C ompare these values with the manometer and
note value.
• C ompare these values with the pressure
reference. The difference between the displayed
pressure reference on the pump and that read on
the manometer should not be more than 5% of
the pressure reference.
Note: The value read on the manometer is in mbar.
Use the following formula to convert mbar to feet
H20:
Feet H20 = Value in mbar displayed on the manometer
2.9993
Repeat this test at level 75, 100 and 125.
50

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis