4.4. Pressure Alarms

High pressure alarm
If the high pressure safety
level of 140 is exceeded,
the pump stops and the red
ALARM (6) LED lights up.
The operator should lower
the pressure to let the pump
automatically restart.
If the pump does not restart
when the pressure drops
below 140, restart the pump
manually by turning the pump
OFF and ON again with the
POWER pad.
Note: numbers in ( ), refer to diagram in section 1.5.
Low pressure alarm
If the pressure falls below the
low pressure safety level of
10, the red ALARM LED lights
up, <Pres L> flashes on the
LCD display, and the pump
stops.
Check that the luer-lock and
the filter near the pressure
transducer are in proper
working order.
If both are satisfactory, press
Fill Chamber to stop the alarm
and Run/Stop pad to restart
the unit.
30
Warning:
If water has accidentally
entered the tubing
leading to the pressure
transducer, change the
One Day Set Irrigation
Tube and restart the
system.