Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 152

Suction Flow Rates Test

Equipment required for the test:
• QC test tubing Ref. 284588
• Four-way foot pedal Ref. 284176
• Timer
• 1 litre graduated beaker
Cannula
Pedal not
Suction
activated
Per minute
50–150 ml
Shaver
Pedal not
suction
activated
Per minute
0 ml
The values recorded should be in the range described in diagram above.
Note: All numbers in ( ) refer to the system indicators figure in section 1.5.
Procedure:
• Switch the FMS DUO+ on "I" on the back panel.
• Press the "POWER" pad (4).
• Press "Run/Stop" pad (18).
• Install the QC test tubing Ref. 284588.
• Place the irrigation tube in a 1 litre bottle of water.
• P lace the outflow tube in the empty beaker to
Pedal
Min
activated
200 ml/mn
Cannula
150–250 ml
Suction
(blue pedal)
(30)
Pedal
Min
activated
200 ml/mn
Black pedal
150–250 ml
47
measure amount of liquid produced by suction for
1 minute.
Low
Med
300 ml/mn
500 ml/mn
225–375 ml
425–575 ml
Low
Med
400 ml/mn
600 ml/mn
325–475 ml
525–675 ml
High
600 ml/mn
525–675 ml
High
800 ml/mn
725–875 ml