Herunterladen Inhalt Inhalt Diese Seite drucken

Suction Settings; Cannula Suction; Shaver Suction; Lavage Mode - DePuy Mitek FMS DUO+ Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 152

4.5. Suction Settings

(Min, Low, Med, High)
Press the Cannula Suction (17) or Shaver Suction
(14) pads to increase the preset suction levels. The
adjacent green indicators will light up to show the
actual suction setting (Min, Low, Med, High).
A - Cannula Suction:
This suction mode is activated when the blue
CANNULA SUCTION (31) foot pedal is pressed.
The adjacent indicator light setting will blink when
the cannula suction is activated as long as you
press on the foot pedal.
B - Shaver Suction:
Step on the black FORWARD (30) or REVERSE
(29) foot pedal to activate your shaver with your
preset suction level. The adjacent indicator light will
blink when the shaver suction is activated.
Note: numbers in ( ), refer to diagram in section 1.5.

4.6. Lavage Mode

(red pedal)
The Lavage mode is activated by:
- Pressing the Lavage pad (10) on the front panel
once.
- Stepping on the red LAVAGE (32) foot pedal for
1.5 seconds. The LED will blink until the lavage
cycle is completed.
To cancel an unwanted or lengthy lavage cycle,
simply step on the blue CANNULA SUCTION foot
pedal (31) or press the Lavage pad (10) on the
front panel.
The shaver can be activated while the lavage cycle
is on.
One beep will sound when the lavage mode is
activated. Two beeps will sound when the Lavage
mode is completed or cancelled.
Caution:
The Lavage mode should only be activated
if an outflow cannula or working cannula is
used. (Use 4.5mm cannula ref. 281580).