Herunterladen Inhalt Inhalt Diese Seite drucken

Electromagnetic Immunity - Acteon satelec Piezotome Solo Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 52

13. 2 ELECTROMAGNETIC IMMUNITY

The device is intended for use in the electromagnetic environment speciÀ ed in the table below. The user and/
or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
IEC 60601 test level
Electrostatic
± 6 KV contact
discharge (ESD)
± 8 KV air
IEC 61000-4-2.
Electrical fast
± 2KV for power
transients
supply lines
IEC 61000-4-4.
± 1KV differential
Surges
mode
IEC 61000-4-5.
± 2KV common mode
<5% U
U
) for 0.5 cycles.
T
Voltage
40% U
dips, short
for 5 cycles
interruptions
and voltage
70% U
variations
for 25 cycles
IEC 61000-4-11.
<5% U
U
) for 250 cycles
T
16
Compliance level
± 6 KV contact
± 8 KV air
± 2KV for power
supply lines
± 1KV differential
mode
± 2KV common mode
(>95% dip in
<5% U
T
U
) for 0.5 cycles.
T
(60% dip in U
)
40% U
T
T
for 5 cycles
(30% dip in U
)
70% U
T
T
for 25 cycles
(>95% dip in
<5% U
T
U
) for 250 cycles
T
Electromagnetic environment - guidance
Floors must be wood, concrete, cement or tiled.
If Á oors are covered with synthetic material
(carpet, etc.), the relative humidity must be at
least 30%.
Mains power quality should be that of a typical
commercial or hospital environment (hospital,
clinic).
Mains power quality should be that of a typical
commercial or hospital environment.
(>95% dip in
T
Mains power quality should be that of a typical
(60% dip in U
)
T
T
commercial or hospital environment.
If the use of the device requires continued
operation during power mains interruption, it is
(30% dip in U
)
recommended that the product be powered from a
T
T
separate power supply (UPS, etc.).
(>95% dip in
T

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis