SYSTEM TECHNICAL DATA

POWER SUPPLY CHARACTERISTICS
TYPE OF POWER SUPPLY
NOMINAL VOLTAGE
MAXIMUM VOLTAGE VARIATION
NOMINAL CURRENT
FREQUENCY
ABSORBED POWER
APPARENT LINE RESISTANCE
PROTECTION FUSES (F1-F2-F3-F4)
CIRCUIT PROTECTION FUSES
TECHNICAL DATA OF THE RADIOGRAPHIC SYSTEM
NOMINAL HIGH VOLTAGE
NOMINAL CURRENT
NOMINAL ELECTRIC POWER AT 0.1 sec
REFERENCE CURRENT-TIME PRODUCT
LOAD FACTOR COMBINATION
INTENSITY OF RADIATION IN THE AIR
TOTAL FILTRATION
HALF VALUE LAYER (HVL) AT 70 kV
LEAKAGE RADIATION
LINEARITY
REPRODUCIBILITY
ELECTRICAL CLASSIFICATION
MEASUREMENT CONDITIONS
kVp
mA
EXPOSURE TIME (sec)
24
CHAPTER 9
Single phase alternate
220 V - 230 V - 240 V
-5% +10%
3.5 A - 3.2 A - 3 A
50/60 Hz
0.8 kVA
0.5 Ω
6.3AF - 250V
(F5) - no. 1 630 mA - 125 V
(F6) - no. 1 500 mA - 125 V
70 kV
8 mA
560 W (70 kV - 8 mA)
0.8 mAs - 8 mA - 0.1 sec
70 kV - 8 mA
38 uGy/mAs ±20% at 1 meter away from focal spot
2 mm Al at 70 kV
2 mm Al
< 0.25 mGy/h to 1 meter away from focal spot
10%
0.05
Class I - Type B - intermittent service
Non-invasive measurement with delay time = 0.05 sec
Direct measurement with digital instrument
Non-invasive measurement to nominal voltage
115 V
-5% +10%
5.5 A
50/60 Hz
0.8 kVA
0.2 Ω
8AF - 250V