2.3
Removing the upper side cover
✔
The back panel is completely removed, see "Removing the back pan-
el" [ ➙ 5].
✔
The front panel is completely removed, see "Removing the front pan-
el" [ ➙ 6].
NOTICE
Pay attention to the screw threads when removing and re-inserting the
screws. The middle screw (B2) is an M4x10 machine screw; the other
two screws (B1) are self-tapping screws.
1. Remove the three screws (B) of the upper side cover. Be careful not
to lose the washer of the middle screw.
2. Remove the upper side cover from the unit sideways.
Removing the covers
Removing the upper side cover