6 Operating
6
Operating
6.1
Preparing the Exposure
4
24
A
B
A
A
11
Loading the Film Cassette
Treat the film cassette and intensifier screen with care to
avoid scratching the screen or denting the cassette
housing.
Opening the cassette: Push the locks A forward as
shown and lift up lever B.
Then you can place film in the cassette.
Closing the cassette: Press the cover down uniformly
on both sides until the locks click.
Insert Accessories
•
Insert Chin Rest or Bite Block up to stop.
For application see chapter "Exposure Programs".
•
Press button (11) to remove.
•
Open Temple Supports with button (4)
•
Insert Forehead Support Strip A
with attachment A ↑ up for adults.
•
with attachment A ↓ down for patients with smaller
•
heads.
Contact Spacer and Head Positioner
•
Insert Contact Spacer L
Application for temporomandibular jaw exposure pro-
grams P4 – P9.
Always place the contact spacer where, as a result of
anatomical features, with correct head positioning there
is no contact with the forehead (no 3-point fixing).
•
After positioning the head, press the contact spacer
onto the tube and push it down to the forehead con-
tact.
•
Insert Head Positioner J
•
Open temple supports,
•
Insert head positioner up to stop.
Application for temporomandibular jaw exposure pro-
grams P4–P9 and for paranasal tomography P13, but
without ear fixation.