• Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning
agent neodisher ® mediclean forte (Dr. Weigert GmbH & Co.
KG, Hamburg, Germany) with the recommended washing
program.
5.4

Recommended methods

Erbe recommends the preparation procedures described below.
Equivalent different procedures are possible if not explicitly ex-
cluded. It is incumbent on the user to ensure the suitability of the
actual procedures used by means of suitable measures (e.g. val-
idation, routine monitoring, check of material compatibility).
5.5

Dismantling

The product must be prepared while it is dismantled.
1. Disconnect the connection cable from the handle (no pic-
ture).
3.
5.
4.
2.
2. Coagulation forceps: Close the handpiece.
3. Press the release button and pull the insert out of the shaft
tube.
21