Instructions for Use

Operating Procedures

Operating Procedures

Operating Modes

The Sublimation Controller can be operated in the MANUAL,
AUTOMATIC, REMOTE SET and AUTOMAT/REMOTE mode.
In MANUAL and AUTOMATIC modes all the controls are given on
front panel.
In AUTOMAT/REMOTE and REMOTE SET, the controls are given via
signals coming through the I/O cable.
After a power failure the Sublimation Controller will return to the previously
NOTE
selected operating mode. If "AUTOSTART" parameter was set on "YES" and
before the power failure the unit was in sublimation mode, the sublimation
starts on again automatically.
The unit is designed to operate in two different modes:
The Sublimation Controller can be used to control the TSP or Mini Ti-
Ball pumps type.
In MANUAL and REMOTE SET operating modes the sublimation
parameters are setting by the user by means of the front panel
controls (MANUAL mode) and/or by signal send on the I/O connector
(REMOTE mode).
If the Sublimation Controller is used in conjunction with the TSP
pump the TSP source type must be set. If the Sublimation Controller
is used in conjunction with the Mini Ti-Ball pump the Mini Ti-Ball
source type must be set.
70/127
with manual selection of the parameters (MANUAL and
REMOTE SET modes)
with automatic selection of the parameters (AUTOMATIC and
AUTOMAT/REMOTE modes)