Tsp controller

Werbung

The communication port mating connector is supplied with the unit.
See the host computer's instruction manual for further details on
serial port connections.
The following picture shows the right procedure to connect a cable to
the I/O or to the serial port connector. A shielded cable of 30 m
maximum length has to be utilized for both serial and I/O port
connections.
Take care to have a good contact (soldered) between the metallic connector
NOTE
case and the external shield of the cable. Moreover, this connection has to be
assured at least on the controller side.
In this way, you will be sure to reduce the influence of the external
noise and to accomplish the EMC requests.
TSP Controller User Manual
SHIELD
ASSEMBLED SHELL
Figure 11
Cable assembling instruction
/
87-900-138-01
Technical Information
Sublimation Controller Outline
Picture A
SHIELD SOLDERED TO
CONNECTOR METALLIC
CASE
Picture B
SHELL
Picture C
5
85/127