5
Technical Information
Use
In this operating mode only the sublimation current value and the
pressure threshold are set by means of the control panel. The
sublimation occurs when the pressure threshold is reached (analog
input on pin 2 of the I/O connector, range 0 to 5 Vdc).
If the ▼ key is pressed the display shows the previous message
(mode MANUAL).
If the ▲ key is pressed the display shows:
M
In this operating mode only the time and period of sublimation are
set by means of the control panel. The current value of sublimation is
set by means of a signal incoming on the analog input on pin 2 of the
I/O connector, range 0 to 10 Vdc (0 Vdc = 0 A, 10 Vdc = 50 A).
If the ▼ key is pressed the display shows the previous message
(mode AUTOMATIC).
If the ▲ key is pressed the display shows:
M
In this operating mode the sublimation starts when the pressure
threshold is reached (as in the AUTOMATIC operating mode).
The current value of sublimation is set by means of a signal incoming
on the analog input on pin 2 of the I/O connector, range 0 to 10 Vdc.
If the ▼ key is pressed the display shows the previous message
(mode REMOTE SET).
If the ▲ key is pressed the display shows the first message illustrated
before (mode MANUAL).
94/127
O
D
E
:
R
E
M
O
D
E
:
A
U
T
O
M
O
T
E
S
A
T
/
R
E
TSP Controller User Manual / 87-900-138-01