Herunterladen Diese Seite drucken

Agilent Technologies 929-0032 Bedienungshandbuch Seite 111

Tsp controller

Werbung

COMMAND
DESCRIPTION
N
Sublimation current
(measurement unit: 0,1 A;
this value is rounded to 0,5
A)
P
Sublimation period
R
Recover
S
Controller status
T
Sublimation time
(measurement unit: 0,1
min.)
V
Absorbed voltage
(measurement unit: 0,1 V)
TSP Controller User Manual
DATA TYPE
Numeric
Numeric
Logic
Numeric
Numeric
Numeric
/
87-900-138-01
Technical Information
Letter Protocol Description
READ/WRITE
ADMITTED VALUES
R/W
300 = 30 A
305 = 30,5 A
......
495 = 49,5 A
500 = 50 A
R/W
30
100
300
600
1200
2400
48000
19200
R/W
0 = automatic
1 = manual
R
0 = stop
1 = fail
2 = wait interlock
3 = ramp
4 = wait sublimation
5 = sublimation
R/W
10 = 1 minute
15 = 1,5 minutes
.....
150 = 15 minutes
R
5
= 3 minutes
= 10 minutes
= 30 minutes
= 1 hour
= 2 hours
= 4 hours
= 8 hours
= 32 hours
111/127

Werbung

loading