4
Instructions for Use
Preparation for Installation
1
2
If the line voltage pass from 220-240 Vac to 100-120 Vac, or viceversa, the power
CAUTION!
cord and the fuses MUST be changed also. For 220-240 Vac the fuses must be
T8A and for 100-120 Vac must be T15A.
3
4
64/127
Switch off the Sublimation Controller and unplug the power cord
from the controller.
On the rear panel slide the voltage selector till the desired
position.
If necessary, change the fuses and the power cord. If power cord
must be changed, rewire as follows:
blue = neutral
brown = phase
yellow-green = ground
Plug the power cord to the controller
Figure 2
Rating voltage selector