For these parameters the possible settings and the messages
displayed are the same shown for the MANUAL and AUTOMATIC
operating mode, excepted for the script A/R in place of MAN at the
end of the second row.
Parameters Values
In the following table are indicated the different possible values for
each sublimation parameter.
Tab. 4
Parameter
Current
Time
Period
Pressure threshold
Sublimation
To start the sublimation proceed as follows:
1
2
TSP Controller User Manual
Default value
30 A
1 minute
3 minutes
1x10
mbar
-7
On the rear panel, turn the power switch on (I position).
Wait for the following messages displayed in sequence:
A
/
87-900-138-01
Adjustable range
30 A to 50 A at step of 0.5 A
1 minute to 7 minutes at step of 30 seconds
CONTINUOUS or
3 minutes to 32 hours with following
intermediate values:
10 minutes
30 minutes
1 hour
2 hours
4 hours
8 hours
1x10
mbar to 1x10
-10
U
T
O
T
E
Technical Information
Use
mbar at step of 1x10
-4