Herunterladen Inhalt Inhalt Diese Seite drucken

Storage; Product Specification; Certificates; Specifications - Mettler Toledo InPro 5000 Bedienungsanleitung

Co2 sensor
Vorschau ausblenden Andere Handbücher für InPro 5000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

52002465_F_IM_InPro5000i_edf_2022_09.qxp_52002465_F_IM_InPro5000_edf_2022
18
7

Storage

For storage we recommend to fill the membrane
body with fresh CO
sensor. The wetting cap, filled with CO
prevent a drying-out of the membrane.
8
Product specification
8.1
Certificates
Each sensor is delivered with a set of 3.1 B certificates
(complying with EN 10204/3.1 B).
All wetted metal parts (sensor shaft, cap sleeve and
membrane body) are identified with an engraved sym-
bol corresponding to the heat number on the paper
certificate delivered with the sensor.
Each wetted metal part (sensor shaft, cap sleeve and
membrane body) is polished in order to get a surface
roughness lower than 0.4 μm (16 μin). This represents
a roughness grade number of N5 (according to ISO
1320:1992).
8.2
Specifications
Measurement principle
Working conditions
Mechanical pressure resistance
Pressure range measurement
Temperature range (measurement)
Sterilization
Construction
Temperature signal
Cable
O-ring material
Membrane material
Wetted metal parts
Dimensions
Sensor diameter
Immersion length
Performance
Measuring range
Accuracy
Detection limit
Response time
InPro 5000 (i)
52 002 465 F
InPro 5000 (i) CO
2
electrolyte and rebuilt the
2
-electrolyte, will
2
InPro 5000 (i)
Potentiometric / Severinghaus
max. 3 bar at 25 °C
[max. 42 psi at 77 °F]
0.2 – 2 bar absolute
[3 – 30 psi absolute]
0...60 °C [32...140 °F]
< 130°C [266 °F] in situ
or autoclave
InPro 5000 i: Digital
InPro 5000: Pt 1000
InPro 5000 i: AK9
InPro 5000: VP
Viton
®
, Silicone (FDA approved)
PTFE-silicone (reinforced)
Stainless steel DIN 1.4435
similar to AISI 316 L
12 mm (0.47")
120, 220, 320 mm
(4.7, 8.66, 12.6")
10 – 1000 hPa CO
± 10 %, 600 – 900 hPa
≥ ± 10 %, > 900 hPa
10 hPa CO
2
90 % of final value < 120 sec
(at 25 °C [77 °F] from air to CO
© 09 / 2022 METTLER TOLEDO
Printed in Switzerland
Sensor 12 mm
2
)
2

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Inpro 5000i

Inhaltsverzeichnis