18
7

Storage

h
For storage we recommend to fill the membra-
ne body with fresh CO
sensor. The wetting cap, filled with CO
prevent a drying-out of the membrane.
8
Product specification
8.1

Certificates

Each sensor is delivered with a set of 3.1 certificates
(complying with EN 10204/3.1).
All wetted metal parts (sensor shaft, cap sleeve and mem-
brane body) are identified with an
sponding to
delivered with the sensor.
Each wetted metal part (sensor shaft, cap sleeve and
membrane b
ness lower than 0.4 µm (16 µin). This represents a rough-
ness grade number of N5 (according to ISO 1320:1992).
8.2

Specifications

Measurement principle
Working conditions
Mechanical pressure resistance max. 3 bar at 25 °C
Pressure range measurement
Temperature range
(measurement)
Sterilization
Construction
Temperature signal
Cable
O-ring material
Membrane material
Wetted metal parts
Dimensions
Sensor diameter
Immersion length
Performance
Measuring range
Accuracy
Detection limit
Response time
InPro 5000 (i)
52 002 465 G
InPro 5000 (i) CO
the heat number on the paper certificate
ody) is polished in order to get a surface rough-
electrolyte and rebuilt the
2
engraved
InPro 5000 (i)
Potentiometric / Severinghaus
[max. 42 psi at 77 °F]
0.2 – 2 bar absolute
[3 – 30 psi absolute]
0...60 °C [32...140 °F]
< 130 °C [266 °F] in situ
or autoclave
InPro 5000 i: Digital
InPro 5000: Pt 1000
InPro 5000 i: AK9
InPro 5000: VP
Viton
, Silicone (FDA approved)
®
PTFE-silicone (reinforced)
Stainless steel DIN 1.4435
similar to AISI 316 L
12 mm (0.47")
120, 220, 320 mm
(4.7, 8.66, 12.6")
10 – 1000 hPa CO
± 10 %, 600 – 900 hPa
≥ ± 10 %, > 900 hPa
10 hPa CO
2
90 % of final value < 120 sec
(at 25 °C [77 °F] from air to CO
© 03 / 2023 METTLER TOLEDO
Printed in Switzerland
Sensor 12 mm
2
-electrolyte, will
2
symbol corre-
2
)
2