26
8.2

Specifications

Measurement principle
Working conditions
Permissible pressure range
during measurement
Mechanical pressure resistance
Permissible temperature
range during measuring
Mechanical temperature
resistance of the sensor shaft
Sensor performance
Operating range
Accuracy in aqueous media
Response time at
25 °C (77 °F) (air N
Design features
Temperature compensation
Cable connection (digital)
Wetted O-rings
O
selective membrane material
2
Wetted sensor parts
Surface roughness of
wetted s / steel parts
(EN 1320: 1996)
Certificates (MaxCert™)
Quality (final inspection certificate)
FDA / USP Class VI
Material certificate 3.1
Surface finish certificate 2.1
InPro 6000 optical O
52 206 256
InPro 6000 Series Optical O
)
2
sensors
2
InPro 6860i/6870 i / 6880 i /
6960i/6970 i
optical
InPro 6860 i / 6880 i / 6870 i:
0.2 ... 6 bar absolute
2.9 ... 87 psi absolute
InPro 6960 i:
0.2 ... 9 bar absolute
2.9 ... 130.5 psi absolute
InPro 6970 i:
0.2 ... 12 bar absolute
2.9 ... 174 psi absolute
InPro 6860 i / 6880 i / 6870 i:
max. 6 bar absolute
max. 87 psi absolute
InPro 6960 i / 6970 i:
12 bar absolute
174 psi absolute
InPro 6860 i /
InPro 6870 i /
InPro 6880 i:
0 ... 60 °C
32 ... 140 °F
InPro 6860 i /
InPro 6880 i:
– 20...140 °C
– 4...266 °F
steam-sterilizable – 4...250 °F
autoclavable
InPro 6880 i shaft
only
InPro 6860 i / 6880 i /
6870 i / 6960 i:
8 ppb to 60 % O
InPro 6970 i:
2 ppb to 2000 ppb
InPro 6860 i / 6880 i /
6870 i / 6960 i:
† ±[1%+8 ppb]
InPro 6970 i:
† ±[1%+2 ppb]
t
< 20 s, < 70 s (InPro 6860 i)
98%
automatic with built-in RTD
data cable 5 pin,VP8 (InPro 6860 i)
EPDM FDA approved (other
materials on request)
silicone, PTFE (InPro 6860 i)
s / steel 1.4404, [AISI 316L] PPS
with material certificate 3.1 (other
materials on request)
Ra < 0.4 µm [16 µin]
yes
yes
yes
yes
© 07 / 12 Mettler-Toledo AG
Printed in Switzerland
Sensors 12 mm
2
InPro 6960 i /
InPro 6970 i:
0 ... 40 °C
32 ... 104 °F
InPro 6870 i /
InPro 6960 i /
InPro 6970 i:
– 20...121°C
steam-sterilizable
saturation
2

Werbung