Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance; Inspection Of The Sensor; Visual Inspection - Mettler Toledo InPro 6950 Serie Bedienungsanleitung

O2-sensoren
Vorschau ausblenden Andere Handbücher für InPro 6950 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

InPro 6950 Series O
6

Maintenance

6.1

Inspection of the sensor

6.1.1 Visual inspection

To check your sensor, we recommend the following
procedure:
– T he contacts of the connector must be dry.
Moisture, corrosion and dirt in the connector can
lead to false readings.
– C heck the cable for buckling, brittle areas or
ruptures.
– B efore calibration always examine the membrane
foil optically for signs of damage. The foil must
be intact and clean. Dirty membranes should be
wip ed clean using a soft, moist tissue.
h
Note: An undulated membrane has no influence on
the sensor performance, assuming the membrane is
intact.
– T he membrane body must be replaced if the
sensor has too long a response time, the reading
is unstable or subject to drift, and if the sensor
cannot be calibrated or the membrane shows sign
of mechanical damage.
– C heck the cathode area for discoloration, conta-
mination or cracks in the glass. If necessary rinse
with demineralized water and clean with a clean
soft brush or soft paper tissue.
Attention! Do not use any cleaning agents contai-
ning alcohol. This could damage the sensor or lead
to fault current.
Attention! The glass body is fragile and sensitive to
vibration.
© 02 / 2023 METTLER TOLEDO