Tests and Approvals
CAUTION
!
The zero point and sensitivity must be tested daily before
every use. The sensitivity must be tested using a known
methane concentration in the range of 25 to 50 % of the full
scale value. The measurement accuracy must be in the range
between –20 % and +20 % (–5 % and +5 % for use in coal
mining) of the specified gas concentration.
Any rapid up-scale reading followed by a declining or erratic
reading may indicate a gas concentration beyond upper scale
limit which may be hazardous.
High off-scale readings may indicate an explosive concentration.
Power Packs
Three different types of power supply unit:
Dräger X-am 7000 NiMH Battery Pack 4.8 V / 3.0 Ah
Dräger X-am 7000 NiMH Battery Pack nominal voltage 4.8 V / 6.0 Ah
Dräger X-am 7000 Alkaline Battery Pack 6.0 V
NiMH Battery Packs
The operating temperature range of an instrument equipped with
a NiMH battery pack is between –20 °C and +60 °C.
Changing the NiMH Battery Packs inside a hazardous location is
allowed.
Do not charge in a hazardous location.
Alkaline Battery Packs
The operating temperature range of an instrument equipped with a
alkaline battery pack is between –20 °C and +40 °C.
38
WARNING
!
Explosion hazard! To reduce the risk of ignition of a flammable
or explosive atmosphere, strictly adhere to the following
warnings:
Do not mix new batteries with used batteries, and do not
mix batteries from different manufacturers or of different
types.
Remove batteries before maintenance work.
Do not replace or open alkaline power packs in a
hazardous area.
Do not replace the alkaline batteries in a hazardous area.
Only use one of the following battery types (size C, round
cell or LR14): Energizer, E93 or Duracell Plus, MN 1400.