Switching On The Instrument - Dräger X-am 7000 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für X-am 7000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Screen 3
The full-scale values of the measuring
ranges for all active sensors.
Screen 4
Information about the owner or user of the
instrument, together with its application
area and the date and time.
If "xx.xx.xx xx:xx" is displayed instead of the date and time, or if
the date and time are incorrect:
(possible only if the clock has been reset as the result of a power
failure)
Set the date and time, see the "Dräger X-am 7000 Technical Manual".
NOTICE
i
i
If the date and time are not set correctly, some functions
(such as calibration) are not possible.
Dräger X-am 7000
4.4
CO
5.0
Vol.%
2
CH
100.0
%LEL
4
CO
500.0
ppm
H
S
100.0
ppm
2
O
25.0
Vol.%
2
3/3
Measuring range
Firma XYZ
Group 123
:
Field
:
34
Tag
:
2M073Ø487A
1/3
Identification
Phase 1 of the power-on sequence:

Switching on the instrument

{ress and hold the » OK « key (longer than 3 seconds).
A count-down is displayed to indicate
when the three second period has
elapsed:
When the countdown reaches zero, release the » OK « key; the
instrument then switches itself on in three phases.
If you release the » OK « key before the countdown reaches zero,
the instrument remains in the off state.
Product information and a message that
the self-test is running are displayed.
During this phase, a signal tone sounds
briefly.
The red alarm lamp lights briefly (test of
the alarm signals).
The green operating LED is switched on.
Operational characteristics
D
X – am 7000
_ 3 _
D
X – am 7000
_ 2 _
D
X – am 7000
_ 1 _
D
X – am 7000
_ _ _ starting selftest _ _ _
41