Operation

7.1

Switching device on and off

7.2

Operation of the Vario footswitch with pedal

1. Use bail to replace the pedal by foot.
2. Foot switch "Vacuum" (blue) starts the vacuum pump.
3. Foot switch "Motor" (yellow) starts the electronic motor to conduct the
lifting movement of the cannula at the Conform Cannula Handpiece.
4. Pedal for starting the infiltration pump and regulating the amount of
infiltration liquid delivery up to the preset maximum value.
7.3

Regulation of suction performance

Establishing airtightness of the suction system and checking for leaks:
1. Turn Vacuum Controller counter clockwise all the way to the stop (equivalent to vacuum minimum).
2. Switch on main switch (I/O) of the pump and press On/Off pedal. Pump is running and building up vacuum.
3. Crimp suchtion tube to generate maximal air-tightness of the suction system.
4. Turn Vacuum Controler clockwise (equivalent to vacuum maximum).
5. Wait for maximum build up of vacuum (equivalent to ≥ -0.9 bar).
6. Now by turning the Vacuum Controller the suction performance can be regulated steplessly from 0 to -0.9 bar.
The vacuum pump is always running at maximum speed. The reduction of the suction performance is
achieved by forced ventilation.
7.4

Regulation of infiltration performance

10
Use the power switch "I/O", at the rear of the device, to switch the device on and off.
The standby is signalized by the LED status light at front of the device.
The device can be switched off at any time, irrespective of any procedure for device
switch-off.
The suction performance is regulated by the Vacuum Controller at the front side of the de-
vice.
Turning clockwise:
Turning counterclockwise:
The Infiltration Controller is used to preset the maximum infiltration performance that
can be retrieved by using the pedal. By pressing the pedal infiltration can be started,
regulated and halted.
Turning clockwise:
Turning counterclockwise:
1
Vacuum increases, suction performance is enhanced.
Vacuum decreases, suction performance is reduced.
Delivery of maximal infiltration liquid is increased.
Delivery of maximal infiltration liquid is reduced.
Operating instructions Vacuson 60 LP, Art. No. 31793, V46/13
Vacuson 60 LP
3
4