Vacuson 60 LP
7

Operation

7.1

Switching device on and off

Use the power switch "I/O", at the rear of the device, to switch the device on and off.
The standby is signalized by the LED status light at the front of the device.
The device can be switched off at any time, irrespective of any procedure for device
switch-off.
7.2

Operation of the VARIO footswitch with pedal

1. Foot switch "Vacuum" (blue) starts the vacuum pump.
2. Foot switch "Motor" (yellow) starts the electronic motor to
conduct the lifting movement of the cannula at the Conform
Cannula Handpiece.
3. Pedal for starting the infiltration pump and regulating the
amount of infiltration liquid delivery up to the preset maxi-
mum value.
7.3

Regulation of suction performance

The suction performance is regulated by the Vacuum Controller at the front side of the de-
vice.
Turning clockwise:
Turning counterclockwise:
Establishing airtightness of the suction system and checking for leaks:
1. Turn Vacuum Controller counter clockwise all the way to the stop (equivalent to vacuum minimum).
2. Switch on device (I/O) and press the foot switch "Vacuum" at the VARIO pedal. Pump is running and
vacuum is building up.
3. Crimp suchtion tube to generate maximal air-tightness of the suction system.
4. Turn Vacuum Controler clockwise (equivalent to vacuum maximum).
5. Wait for maximum build up of vacuum (equivalent to ≥ – 0.9 bar (relative)).
6. Now by turning the Vacuum Controller the suction performance can be regulated steplessly from 0 to
– 0.9 bar (relative).
The vacuum pump is always running at maximum speed. The reduction of the suction performance is
achieved by forced ventilation.
The vacuum manometer shows the current vacuum in the device – due to the connection of tubes and
adapters, the effective vacuum at the cannula can deviate from the displayed value.
7.4

Regulation of infiltration performance

The Infiltration Controller is used to preset the maximum infiltration performance that
can be retrieved by using the pedal. By pressing the pedal infiltration can be started,
regulated and halted.
Turning clockwise:
Turning counterclockwise:
Operating Instructions, Vacuson 60 LP, REF 31793, V34/17
Vacuum increases, suction performance is enhanced.
Vacuum decreases, suction performance is reduced.
Delivery of maximal infiltration liquid is increased.
Delivery of maximal infiltration liquid is reduced.
1
3
15
EN