Condensate drain
Compressors (Fig. 14)
Draining the condensate from the pressure vessel is
recommended during regular operation. Switch off the
compressor at the mains. Reduce air pressure in the device
to max. 1 bar by releasing air via a connected device. Install
the hose from the basic package into the drain valve. Place
a vessel under the hose and open the drain valve (1) and
drain the condensate from the tank.
Compressors with condensation and filtration unit
(Fig. 15)
During regular use, condensation is automatically drained
via the drain valve of the condensation unit filter. To check
that the automatic drain is working properly, open the valve
(2) of the drain vessel (1) by turning to the left. Release a
small amount of condensate from the vessel. Close the
valve (2) by turning to the right.
Both compressor types can be equipped with an automatic
condensate drain on the air tank to allow condensate to
drain without operator intervention (see the Accessories
chapter).
Compressors with air dryer (Fig. 16)
Condensate
from
automatically drained into a collecting bottle. The bottle
must be drained regularly, see Chapter 19.1
THE FOLLOWING IS REQUIRED BEFORE ANY FOLLOW-UP CHECKS
For cabinet-mounted compressors, unlock the door and open the door on the cabinet.
Safety valve check
(Fig. 17)
When the compressor is operated for the first time, make sure
that the safety valve is working properly. Turn the screw (2)
on the safety valve (1) several rotations to the left until the
safety valve releases air. Let the safety valve vent for only a
few seconds. Turn the screw (2) to the right until it seats,
closing the valve.