Altana BYK LCS III Betriebsanleitung Seite 69

Vis-spektralphotometer mit einem wellenlängenbereich von 320 bis 1100 nm.
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Standard Operations
Option
More
Store icon
Time & Interval
λ
View Table
Scale & Units
Send Data
Recall Data
Instrument Setup
68
Table 11 Time course setup options
Description
For further Options
To store the scan data
To input the total time for data collection and the time interval between the collection of data
points
To input the wavelength setting
To display readings in absorbance, transmittance or concentration. This can be changed after
sample data are collected
Scale: In the automatic scaling mode, the y-axis is automatically adjusted so that the total scan is
displayed.
The manual scaling mode allows sections of the scan to be displayed.
Units: Choice of absorbance or transmittance.
To send Data to a printer, computer or USB memory stick (Type A)
Call up saved measurement data, wavelength scans or time courses, see
page
28.
Basic data of the instrument, see
Time & interval:
1. Press Time & Interval in the Options menu.
2. Input the total time and the reading time and press OK to
confirm.
Note: In total 500 measuring steps are possible. To select a total time and
a time interval that would cause this number of measurements to be
exceeded, the time interval is defined automatically and the OK key is
inactivated.
section 5.2 on page
17.
section 5.3 on