Standard Operations

5.3 Store, recall, send and delete data

5.3.1 Store, recall, send and delete data from the color log and data log

5.3.1.1 Auto/manual data storage
5.3.1.2 Recall stored data from the color log or the data log
28
The data log of the instrument is seperated in 4 segments:
Color Log, Wavelength Scan, Time Course and Data Log.
The Color Log will store up to 500 color measurments taken in the
Color Measurements and Color Difference Measurement modes.
The Wavelength Scan log will store up to 20 scans.
The Time Course log will store up to 20 time course data sets.
The Data Log will store 500 additional readings in the Photometry
(Single Wavelength and Multi Wavelength mode).
A complete record of the analysis is stored, including the Date,
Time, Results, Sample ID and Operator ID.
The data storage parameter indicates whether data are to be stored
automatically or manually (in which case the user has to decide
which data to store).
1. Press Store: On/Off in the Options menu.
Store On setting: All measurement data are stored
automatically.
Store Off setting: No measurement data are stored.
However, this setting can be changed to Store On in the
result display through Configuration. The reading currently
shown in the display is then stored.
Note: When the instrument memory (data log) is full, the oldest data are
automatically deleted allowing the new data to be stored.
1. Press Recall Data in the Main Menu.
2. Press Color Log or Data Log.
A listing of the stored data is displayed.
3. Press Filter: On/Off.