Herunterladen Inhalt Inhalt Diese Seite drucken

Zero Point Correction (Only For Gauge Pressure Range) (Offset Correction) - LR-Cal LPC 200 Betriebsanleitung

Druckkalibrator
Vorschau ausblenden Andere Handbücher für LPC 200:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Pressure Calibrator
LPC 200
6.6.5 Zero point correction (only for gauge pressure ranges)
(OFFSET CORRECTION)
Different configurations can be carried out via the FUNC key, e.g. the zero point
correction of gauge pressure sensors.
Activating the zero point correction (OFFSET CORRECTION) menu item by:
pressing the FUNC key five times* (directly in the measuring mode).
* If you already are in a "FUNC menu item", press the FUNC key until the "OFFSET CORRECTION" menu
item appears.
Zero point correction for gauge pressure sensors
If the measuring value of the LPC 200 is not zero with the gauge pressure
sensor connected although the sensor is not pressurised and open to atmos-
phere, the zero point of the gauge pressure sensors can be corrected via the
"OFFSET CORRECTION" "FUNC menu item" by pressing the TARE key.
Return to the measuring mode by pressing this key several times.
FUNC
LR-Cal Operating Instructions
Pressure Calibrator LPC 200
GB
7

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis