Pressure Calibrator
LPC 200

2. Charging the internal battery

GB
The mains plug of the battery charger unit powering the LPC 200 must always be
accessible when connected to a power outlet, i.e., you must be able to pull the plug
from the power outlet without difficulty, at any time. However if possible, for safety
reasons, it should be operated without the battery charger attached.
The equipment is delivered with the batteries 5 % - 50 %charged and they should
first be fully charged before initial operation.
The battery level status (charge in %) is displayed on the top right side of the
display (see chapter 4.1: "Display and keyboard explanation").
The temperature during charging must be between 10 °C and 45 °C.
Only use the battery charger approved by the manufacturer (LR-Cal by Leitenberger).
When the battery charger is not in use, its power supply plug should be discon-
nected from the mains socket. Do not leave the battery charger attached to
the rechargeable battery longer than one day, since overloading can shorten
its lifespan. If, after 4 hours, the rechargeable battery isn't fully charged, you
should contact the manufacturer. When unused, a fully charged battery will lose
its charge over time.
10
The integrated rechargeable battery may only be charged
outside the area subject to explosion hazards. This means,
connecting the rechargeable battery charging device inside
the area subject to explosion hazards is strictly prohibited.
Use with a defective power supply unit (e.g. short-circuit
from mains voltage to the output voltage) can produce lethal
voltages within the equipment!
When the mains lead/battery charger is connected to the
LPC 200, the rechargeable battery will be charged, even if the
LPC 200 is turned off.
The battery level during storage or shipping should be
between 30 % - 50 %.
LR-Cal Operating Instructions
Pressure Calibrator LPC 200