Herunterladen Inhalt Inhalt Diese Seite drucken

General Cleaning Guidelines; General Notes - Medela Thopaz+ Gebrauchsanweisung

Cardiothoracic drainage
Vorschau ausblenden Andere Handbücher für Thopaz+:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 51
Error
Problem
number
description
Internal error
XXX
Safety related tip
The acoustic alarm «Leak in system – 301» is deactivated for approximately 6 minutes
after Thopaz + is switched on.

13 General cleaning guidelines

WARNINGS
After each use, the parts that have been in contact with aspirated secretions are to be
cleaned and disinfected or disposed.

General notes

– For detailed information refer to Medela cleaning guideline: 200.2391.
– This cleaning guideline relates to the cleaning of the entire Thopaz + product ranges see
Appendix A.
– Never place electrical devices in water or other liquids, specifically Thopaz + and the docking
station.
– Do not spray or pour liquid directly onto Thopaz + /docking station.
– Thopaz + product range cannot be sterilized.
– Rinsing the Thopaz + product range in a washing machine is not permitted.
– Immersion disinfection, thermal disinfection and ultrasound cleaning for the Thopaz + product
range are not permitted.
– Higher temperature than 45°C, 113°F can cause protein coagulation which can lead to
problems further in the process.
– Clean all surfaces immediately after use to avoid residues from drying and to prevent growth
of microorganisms.
– Do not use cleaning agent/detergent based on phenol.
– Do not use steel brushes or steel wool for cleaning.
– Store medical products dry and dust free.
42 |
Troubleshooting and alarms | General cleaning guidelines
Visible on display
Troubleshooting
instructions
1. Switch Thopaz + off
and on again
2. Inform Medela
Customer Service
Remarks / Potential
source of error
Pressure

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis