7.6
Check therapy progress
7.6.1
Main display
Basic display
Advanced display
28 |
Preparation for use / operating instructions
Two different main displays are available.
In the basic display, the current size of the air leak (ml/min) is
shown on the left side. The total amount of collected fluid (ml)
during the therapy is shown on the right.
Press «Advanced» to switch to the advanced display.
In the advanced display, the current size of the air leak (ml/min)
is shown on the left side. Additionally an air leak history of 12 h/
100 ml/min is displayed. For greater air leaks use the air leak
history (see chapter 7.6.2 – History). The total amount of
collected fluid (ml) during the therapy is shown on the right.
Additionally, two customized fluid parameters (ml) are displayed.
To customize these fluid parameters refer to chapter 10 –
changing settings.
Press «Basic» to switch to the basic display.
– Air leaks between 0 and 1000 ml/min are displayed in 10 ml
steps.
– Air leaks over 1000 ml/min are displayed in 100 ml steps.
– The fluid measurements are displayed as accurate as follow:
In a 0.3 l canister + /- 1 ml, in a 0.8 l canister + /- 5 ml and in a
2 l canister + /- 10 ml.