Herunterladen Inhalt Inhalt Diese Seite drucken

Carry Out Functional Check - Medela Thopaz+ Gebrauchsanweisung

Cardiothoracic drainage
Vorschau ausblenden Andere Handbücher für Thopaz+:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 51
7.5
Carry out functional
check
26 |
Preparation for use / operating instructions
1.
2.
3.
4.
3 sec.
1.1 Seal off the conical con-
nector of the patient tub-
ing with your thumb
(through the packaging).
The tubing must not be
kinked otherwise the
measuring tube will be
closed.
2.1 Switch on pressure by
pressing «on» – pressure
is built up.
3.1 Check air leak value.
Air leak decreases:
Thopaz + is ready for use.
Air leak does not decrease:
Leak in the system:
– check tubing connection.
– check seal is correctly
placed.
– check pressure relief valve
for tight connection.
– Replace canister and/or
tubing if necessary.
4.1 Switch Thopaz + to
Standby by pressing
«Standby» for longer than
3 seconds.