7.7

Catheter check

Important: the catheter check works only with a patient
catheter (drain) in the interpleural space.
– The catheter check is only active if the air leak is 0 ml/min.
– Follow instructions on screen.
– A pressure difference from minimal 3 cmH
inspiration [
] and expiration [
leak of 0 ml/min. This pressure difference is an indication for
the respiration cycle of the patient and confirms that the
patient catheter (drain) is not clogged.
Preparation for use / operating instructions
Press [
] to get into the air
leak history. Scroll to page 4/4
by pressing «Next» [
1
2
O between
2
] is displayed at an air
Patient catheter (drain)
open
Patient catheter (drain)
clogged?
].