Electromagnetic compatibility (EMC, IEC 60601-1-2:2007, Table 1)

Electromagnetic emissions

Thopaz + is intended for use in the electromagnetic environment specified below.
The customer or the user of Thopaz + should assure that they are used in such an
environment.
Emission Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
WARNINGS
Thopaz + should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, Thopaz + should be observed to verify normal operation in the
configuration in which it will be used.
Compliance
Electromagnetic environment – guidance
Thopaz + uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
with nearby electronic equipment.
Thopaz + is suitable for use in all establishments,
Class B
including domestic establishments and those
directly connected to the public low-voltage
Class A
power supply network that supplies buildings
used for domestic purposes.
Complies