Medela Thopaz+ Gebrauchsanweisung Seite 41

Cardiothoracic drainage
Vorschau ausblenden Andere Handbücher für Thopaz+:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 51
Visible on display
Error
Problem
number
description
306
Canister full
401
Battery low
402
USB connection
not permitted
405
Standby
406
Temperature
too high
408
Canister change
detected
409
Fluid level
sensor out
of order
301
Leak in system
302
Tubing clogged
305
Battery empty
311
Selftest failed
313
Filter clogged
Thopaz +
315
overheated
407
Fluid alarm
Troubleshooting
instructions
1. Clamp tube
2. Go into Standby
3. Replace canister
4. Continue with «On»
5. Unclamp tubing
Connect Thopaz +
to mains
Unplug USB cable
Switch Thopaz + on or off In Standby for 5 minutes. No
Prepare backup device
Please confirm if canister
has been changed
– Fluid measurement
unavailable
– Confirm and continue
therapy
Check system for leaks
1. Clamp catheter
2. Clamp tubing
3. Go into Standby
4. Replace tubing
5. Continue with «On»
6. Unclamp catheter
Connect Thopaz + to
mains
Snap the canister out
and
in again
1. Clamp tubing
2. Replace canister
3. Continue with «On»
4. Unclamp tubing
Replace Thopaz +
Patient is losing a large
amount of fluid
Remarks / Potential
source of error
Remaining time of
battery approx. 30 min.
USB connection not
permitted during
operation or in the
vicinity of the patient.
Do not expose Thopaz +
to heat sources.
– Disconnection.
– Missing seal at
vacuum port.
– Tubing is kinked or
clogged for approx.
10min.
– Thopaz + detects
clotting of Thopaz +
tubing, NOT of the
patient catheter.
Do not switch Thopaz +
on if the patient is already
connected.
Filter in canister clogged. No
Troubleshooting and alarms
Pressure
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
No
No
Yes
| 41