Measurement
Mode
range
DC voltage
400 m V
4
V
40
V
400
V
Overload protection : 1000 VDC less than 10s
AC voltage
400 m V
4
V
40
V
400
V
Frequency: 50 Hz to 500 Hz
Input resistance during AC and DC voltage measurement
(except mV range) >10 MOhm
Overload protection :
700 VAC rms less than 10s
Resistance
400
4 k Ω
40 k Ω
400 k Ω
4 M Ω
Capacity
4
nF
40
nF
400
nF
µF
4
µF
40
µF
400
4 m F
Measurem. freq./voltage: 700 Hz (square)/approx. 1.5V
70 Hz (square)/approx. 1.6V from 40 nF
Diode test Ge to GaAs ±(2.0% + 5 digits) 1mV
Test current 1mA max. ; forward voltage 5.0V max.
400 mA
DC current
20
A
Overload protection : see 8.3
AC current
400 m A
20
A
Frequency of AC current : 50 Hz to 500 Hz
40
Accuracy
Resolution
±(0.75%+10dgts)
0,1mV
----------"----------
1 mV
----------"----------
10 mV
----------"----------
100 mV
1
V
±(1.0%+10dgts)
0,1mV
1 mV
---------"---------
---------"---------
10 mV
---------"---------
100 mV
±(1.5%+10dgts)
1
V
±(1 %+5dgts)
0,1 Ω
---------"---------
1
---------"---------
10
---------"---------
100
1 k Ω
---------"---------
10 k Ω
±(3.0%+5dgts)
1 pF
---------"---------
10 pF
---------"---------
100 pF
1 nF
---------"---------
---------"---------
10 nF
±(5.0%+10dgts)
100 nF
1 µF
No data
to 40nF
100 µA
±(1.2%+5dgts)
10 mA
±(1.5%+10dgts)
100 µ A
±(1.5%+10dgts)
10 mA
Measurement
Mode
range
1 kHz
Frequency
100 kHz
Sensitivity equal to or greater than approx. 500 mV
pp from 1 Hz to 100 kHz
Period and
µs
1000
pulse width
measurement
0 to 100%
Overload protection : 250 VAC rms / VDC less than 10s
Relation betw.
period and pulse
0 bis 100%
Cycle Duty
Overload protection : 250 VAC rms / VDC less than 10s
1 kHz
Signal
generator
10 kHz
output SIG-OUT
50 kHz
The TTL level is adjustable from 1 Hz to 20 (19.99) kHz
at 5V fixed amplitude
8.3 Maximum inputs
Current measurement: In the A range, 250 VDC/VAC rms max. and 20 A
AC / DC max. for 30 seconds duration with a sub-
sequent cooling phase of at least 15 minutes;
overload protection 20 A 250 V quick-blow cera-
mic fuse (dimensions: 6 x 30 mm)
In the mA range, 250 VDC/VACrms max. and
500mA AC/DC max.; overload protection 0.5 A
250 V ceramic quick-blow fuse (dimensions: 5 x 20
mm)
Attention!!!
The measurement functions diode test, "SIG OUT" (signal generator
output), logic test and capacity measurement are not protected against
overload or excessive input voltage(s).
Exceeding the maximum permitted input levels can damage the measu-
ring instrument and/or endanger the life of the user.
Accuracy
Resolution
±( 0.1%+5 dgts)
1 Hz
---------"---------
±(3.0%+5dgts)
1 µs
1 ms
±(3.0%+5dgts)
0,1%
±(0.1%+1 dgt)
0,1 Hz
---------"---------
1
Hz
---------"---------
41