Herunterladen Inhalt Inhalt Diese Seite drucken

Data Logger; Manual Recording ("Func-Stor") - WIKA CPH6300 Betriebsanleitung

Hand-held druckmessgerät
Vorschau ausblenden Andere Handbücher für CPH6300:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6 Commissioning, Operation
6.5

Data Logger

The device supports two different logger functions:
Manual recording by keypress "store"
"Func-Stor":
Additional input of measuring point (L-Id) is needed
Automatic recording at intervals of set cycle time
"Func-CYCL":
1-channel version:
The logger stores 3 measuring values per data set:
"CALC off": measuring or average (depends on function), min-peak and max-peak
1)
"CALC active"
: measuring or average (depends on function), Calculated value and dummy
value
2-channel version:
The logger stores 9 measuring values per data set:
measuring or average (depends on function), min-peak and max-peak sensor 1
measuring or average (depends on function), min-peak and max-peak sensor 2
measuring or average (depends on function),
"CALC off": min-peak and max-peak sensor 1 – sensor 2
1)
"CALC active"
: Calculated value and dummy value
Min- and max-peak are the minimum and the maximum of the measured values since the last
recording.
This allows precise analysis of the current pressure as well as pressure fluctuations.
For the evaluation of the data the software GSOFT (version V3.0 or higher) has to be used. The
software also allows easy configuration and starting of the logger. (Free updates are available on
our website www.wika.de)
When the logger is activated (Func Stor or Func CYCL) the hold function is no more available, the
key "STORE" is solely used for the operation of the logger functions.

6.5.1 Manual Recording ("Func-Stor")

a) Save measurements manually:
Up to 1000 measurements can be saved if logger function "Func store" is selected
(see "Configuration"):
Press shortly: data set is saved ("St. XX" is displayed shortly,
Input of measuring point "L-Id": Selection of measuring point via keys
Number 0 ... 19999 or text assigned to number 1 ... 40
Assignment of texts can be done with free software "ConfigSoft" which can be
downloaded on our website www.wika.com.
Confirm input with
1)
CALC-Function: see chapter 6.4.5
WIKA Operating Instruction, Model CPH6300
where XX is the number of the data set)
.
GB
or
.
21

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cph6300-s1Cph6300-s2

Inhaltsverzeichnis