6 Commissioning, Operation
Menu
Parameter Value
GB
Set Configuration: General configurations
14
Description
or
Unit: Select measuring unit
Arrow to bar,
Measuring value is displayed in corresponding unit, the
mbar ...
selectable units depend on connected sensor
Arrow to
Measuring value is displayed in freely adjustable user-unit
"user"
bar, mbar
Unit "user": base unit
0000, 000.0
Unit "user": decimal point setting (only if user-unit selected)
... .0000
-19999 ...
Unit "user": multiplication factor
19999
Sea Level: Sea-level correction
oFF
Sea-level correction off
on
Sea-level correction on (see6.4.3)
-2000 ... 9999 Sea-level in [m]
Rate: Measuring rate
Slo
Slow measuring rate (4 Hz filtered, low power consumption)
FASt
Fast measuring rate, filtered (1000 Hz)
P.dEt
Peak detection: fast measuring rate, unfiltered (1000 Hz)
Averaging Filter
1 ... 120
Averaging period in seconds
oFF
Averaging function is deactivated
Auto Power-Off: Select power-off delay
1 ... 120
Power-off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed/no interface communication takes
place.
oFF
Automatic power-off function deactivated (continuous operation)
Background illumination
oFF
Illumination deactivated
5 ... 120
Turn off illumination after 5 ... 120s
on
Illumination always on
(factory settings: 5 s)
WIKA Operating Instruction, Model CPH6300
*,**
*,**
*,**
*
*