6 Commissioning, Operation
2-channel version:
Storable data sets: 8000
GB
A data set contains at slow measuring rate (rAtE SLo):
Sensor 1: measuring value at the time of the recording
■
Sensor 1: min-peak, max-peak since the last recording
■
Sensor 2: measuring value at the time of the recording
■
Sensor 2: min-peak, max-peak since the last recording
■
Difference sensor 1-sensor 2: measuring value at the time of the recording
■
"CALC off":
■
"CALC aktiv"
A data set contains at fast measuring rate (rAtE FASt und rAtE P.dEt):
Sensor 1: arithmetic average since the last recording
■
Sensor 1: min-peak, max-peak since the last recording
■
Sensor 2: arithmetic average since the last recording
■
Sensor 2: min-peak, max-peak since the last recording
■
Difference sensor 1-sensor 2: arithmetic average since the last recording
■
"CALC off":
■
"CALC aktiv"
At measuring rate "rAtE-Slo" a power saving function is selectable: "Lo.Po". If activated
a measurement is only done if it is also logged (i.e. no additional measurements
between the saved values). This seriously reduces the power consumption and is
therefore especially appropriate for long-term measurements
(e.g. leak testing) without an external power supply.
a) Start recording:
Press for 2 seconds: Start Menu, press
Each storage process is signaled by the short display of 'St.XXXXX', where XXXXX is the number
of the saved data set.
If the logger memory is full, the recording stops automatically and the display shows
If Low-Power-Logger-Function "Lo.Po = on" is activated the device switches itself off as soon as
the memory gets filled.
1)
CALC-Function: see chapter 6.4.5
24
Difference sensor 1-sensor 2: min-peak, max-peak since the last
recording
1)
: Calculated value at the time of the recording and dummy value
Difference sensor 1-sensor 2: min-peak, max-peak since the last
recording
1)
: arithmetic average of calculated value since the last recording and
dummy
again: automatic recording will be started.
WIKA Operating Instruction, Model CPH6300
.