6 Commissioning, Operation
Menu
Parameter
Set Output: Universal output adjustments
GB
Set Corr: Input adjustment (1-channel version)
Set Corr: Input adjustment (2-channel version)
16
Value
Description
or
oFF
Interface and analog output off
SEr:
Serial interface activated
dAC:
Analog output activated
01,11..91
Base address for serial interface communication
(only at Out = SEr)
CH 1, CH 2 or
Channel that should be used for analog output
CH
(only at Out = dAC)
DIF
sensor dep.
Zero-point setting for Out = dAC: Measuring value which
e.g.-5.00..5.00 mbar
should correspond to output 0 V
sensor dep.
Slope setting for Out = dAC: Measuring value which should
e.g.-5.00..5.00 mbar
correspond to output 1 V
Zero adjustment/offset of sensors
No zero adjustment of sensors
oFF
The offset of sensor will be displaced by this value to
sensor dep.
compensate for deviations of the probe or of the measuring
e.g.-5.00..5.00 mbar
device.
Slope adjustment of sensors
oFF
No slope adjustment of sensor
The slope of sensor will be changed by this factor [%] to
-2.000 ... 2.000
compensate deviations of probe or measuring device
Zero adjustment/offset of sensors 1
oFF
No zero adjustment of sensors 1
sensor dep.
The offset of sensor 1 will be displaced by this value to
e.g.-5.00..5.00 mbar
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors 1
oFF
No slope adjustment of sensor 1
The slope of sensor 1 will be changed by this factor [%] to
-2.000 ... 2.000
compensate deviations of probe or measuring device
Zero adjustment/offset of sensors 2
oFF
No zero adjustment of sensors 2
sensor dep.
The offset of sensor 2 will be displaced by this value to
e.g.-5.00..5.00 mbar
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors 2
oFF
No slope adjustment of sensor 2
The slope of sensor 2 will be changed by this factor [%] to
-2.000 ... 2.000
compensate deviations of probe or measuring device
WIKA Operating Instruction, Model CPH6300
***
*
*
***