4. Technical data and application requirements

4. Technical data and application requirements
Pressure limitation
Maximum pressure: 200 kPa (2 bar)
Operating temperature
Ambient: -40 ... +60 °C
Medium: -40 ... +60 °C
Ingress protection
IP 54 per EN 60529 / lEC 529

5. Installation

The inclined tube manometer must be screwed to a suitable vertical
surface
The device must be fastened horizontally using the enclosed self-
tapping screws

6. Commissioning

Filling
Turn the zero knob back as far as it will go, and then about one turn
clockwise, to enable setting to be made in both directions. Unscrew the
'fill' knob and fill it with thegauge filling liquid, until the liquid is visible in
the measuring scale near the zero point. Using the zero knob, set it to
zero exactly and re-seal the filling hole.
Connection to the air filter
Connect the instrument's
and also the instrument's
with hoses. The green and red labels must be stuck next to each other
WIKA Operating instructions air2guide Model A2G-30
connection to the outlet side of the filter
i
connection to the intake side of the filter
j
GB
5