8 Faults

7.4

Recalibration

DKD certificate - Certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with time
intervals of approx. 12 months. Every factory recalibration includes, additionally, an extensive free-
of-charge check of all system parameters with respect to their compliance with the specification.
The basic settings will be corrected if necessary.
8 Faults
Display
Description
No sensor connected
Connected sensor or device
defective
Err.9
or
Value extremely out of measuring
range
Battery empty
No display or
Mains operation: wrong voltage or
confused
polarity
characters,
System error
device does not
react on keypress
Device defective
Measured value above allowable
Err.1
range
Sensor defective
Measured value below allowable
Err.2
range
Sensor defective
Display range overflow
Err.3
Err.4
Display range underflow
Value could not be calculated
Er.11
Calculation overflow happened
System error
Err.7
Sensor not present / recognised
– – – –
could not calculate value
CAUTION!
If faults cannot be eliminated by means of the measures listed above, the instrument
must be shut down immediately, and it must be ensured that pressure and/or signal
are no longer present, and it must be prevented from being inadvertently put back into
service. In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter "9.2 Return".
WIKA Operating Instruction, Model CPH6300
What to do
Switch off device and connect sensor
If 2nd sensor available, check if device is ok.
Return defective device/sensor to
manufacturer for repair
Check: pressure not within sensor range?
Replace battery
Check power supply, replace it when
necessary
Disconnect battery and power supplies, wait
shortly, then reconnect
Return to manufacturer for repair
Check: pressure not within sensor range?
-> measuring value to high!
Return to manufacturer for repair
Check: pressure not within sensor range?
-> measuring value to low!
Return to manufacturer for repair
Check: value above 19999 -> to high to be
displayed
Check: value below -19999 (Tara?) -> to low
Choose different unit
Choose different unit
Return to manufacturer for repair
reconnect sensor,
during logging: stop the logger and restart it
suitable sensor / unit combination necessary
GB
29