4. Design and function
Pos.
Button
7
EN
8
9
4.2

Integrated data logger

The CPG1500 features an integrated data logger.
This data logger can be switched on and set via [MENU] / [Logger] (respectively).
4.3

Voltage supply

Three AA alkaline batteries are used as the voltage supply for the instrument. These are
included in delivery.
The battery life time is up to 2,000 hours for continuous operation (without backlighting
and with WIKA-Wireless deactivated).
In the upper right half of the display there is a symbol for the battery capacity. Instructions
on batteries (see chapter 8.2 "Battery replacement").
16
MENU button
Calling up the menu
By activating the [MENU] button, the menu mode is enabled. If the CPG1500
is already inside the menu mode, depending on the display, "L" or "R" will be
done. If the button is pressed for longer, after 2 seconds the cursor switches
(right ◀ or ▶ left).
Inputs are confirmed with the [MENU] button.
LIGHT button
Turning the backlighting on and off
By activating the [LIGHT] button (short or long press) the light is switched on.
The duration of the light being on depends on "LIGHT-OFF" in "SETTINGS".
1 x pressing the [LIGHT] button (Light = On)
2 x pressing the [LIGHT] button (Light = Off)
If the CPG1500 is in menu mode, with a short press on the [LIGHT] button,
the cursor can be moved downwards.
ZERO button
The current pressure value will be set to "0" (gauge) or reference pressure
(abs.).
By activating the [ZERO] button, the current pressure value is set to "0". A
maximum of 5 % of the measuring span can be corrected.
If the CPG1500 is in menu mode, with a short press on the [ZERO] button, the
cursor can be moved upwards.
For gauge pressure measuring instruments, in the range of the zero point,
±5 % of the measured value will be set to "0".
With absolute pressure sensors, on pressing the [ZERO] button, an input
window appears. Here, the current reference pressure must be entered.
The reference pressure must also be within ±5 % of the initial absolute
pressure of the instrument, then the measured value will be set to the
reference pressure entered.
WIKA operating instructions, model CPG1500