Table 3
Guidance and declaration of manufacturer-electromagnetic immunity
Nowadays, many RF wireless equipments have being used in various healthcare locations where medical equipment and/or systems
are used. When they are used in close proximity to medical equipment and/or systems, the medical equipment and/or systems' basic
safety and essential performance may be affected. Arm-type Fully Automatic Digital Blood Pressure Monitor has been tested with
the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2014+AMD1:2020. The customer and/
or user should help keep a minimum distance between RF wireless communications equipment and this medical equipment and/or
Test frequency (MHz)
385
380-390
450
430-470
710
745
704-787
780
810
870
800-960
930
1720
1945
700-1990
1970
systems as recommended below.
Band (MHz)
Service
a)
TETRA 400
GMRS 460
FRS 460
LTE Band 13, 17
GSM 800/900, TETRA
800,
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3,
4, 25; UMTS
Modulation
a)
Modulation d'impulsion
18Hz
b)
FM
c)
±5 kHz deviation
1 kHz sine
Pulse modulation
217Hz
b)
Pulse modulation
18Hz
b)
Pulse modulation
217Hz
b)
USER MANUAL NEOTENS | WWW.SPENGLER.FR
EN
IMMUNITYTEST LEVEL
(V/M)
27
28
9
28
28
63