The pressure detected exceeds
E03
Error
300 mmHg
E04
Warning
Abnormal measurement data
The inflation speed does not meet
E05
Warning
the standard.
Cannot complete the measurement
E06
Warning
normally
Measurement data exceed the
E07
Warning
standard range
E08
Error
Low battery
The pressure exceeds 15 mmHg for
E09
Warning
more than 3 minutes
E20
Error
Firmware check error
E21
Error
Abnormal internal voltage
E41
Warning
Pressure offset check error
Note: If you cannot solve the abnormal situation by yourself, you can consult the manufacturer or the service desig-
nated by the manufacturer, or you can consult Spengler Technical support center or your retailer. It is forbidden to
disassemble and repair without permission.
52
USER MANUAL NEOTENS | WWW.SPENGLER.FR
Please fasten the cuff again, perform a new measure-
ment and check that the pressure is within the range.
Please check if the cuff is worn properly before re-mea-
suring
Please check if the cuff is worn properly before re-mea-
suring
Please check if the cuff is worn properly before re-mea-
suring
Please readjust the cuff and repeat the measurement.
If the problem is not solved, please contact the man-
ufacturer.
Replace batteries or connect the power adapter (if any).
After the warning , the arm cuff will automatically
deflate and a new measurement can be taken.
Please contact the manufacturer
Please contact the manufacturer
Please contact the manufacturer