5. INOGEN ROVE 6 DESCRIPTION
The Inogen Rove 6 Portable Oxygen Concentrator System may include the following accessories:
AC power supply, DC power cable, rechargeable battery pack and carry bag.
This section is intended to help familiarize you with the device's components and interface.
Do not perform any actions on or with your POC until after reading Section 6, General Instructions
Inogen Rove 6.
Flow setting control buttons
Display
Audible
signals
Power button
• Pressing and holding this button turns the
device on and off.
Flow setting control buttons:
• Use the – or + flow setting control buttons to
change the setting.
• There are six settings, from 1 to 6.
Volume control button:
• Pressing this button will change the volume
level from 1 to 4.
Bell button:
• Pressing this button will toggle the device's
no-breath-detect audible alarm on and off.
º When this mode is ON: The device will alarm
with audible and visual signals when no
breath has been detected for 60 seconds. At
60 seconds, the device will enter 'auto pulse
mode'. Once another breath is detected, the
device will exit 'auto pulse mode' and deliver
normally on inspiration.
º This mode is enabled when there is a bell
"displayed on the screen." If power is lost,
the no-breath-detect audible alarm remains
set in the user preferred mode.
Display:
• The display shows information about the status
of the device such as flow setting, power
status, battery life and alarms.
• Before use, remove the static cling FCC label
from the screen.
©2024 Inogen, Inc. All rights reserved.
Bell
button
Power button
& Backlight
Audio
button
Indicator
lights
Power
in
Indicator lights:
• Breath Detect LED: A green light indicates
breath detection.
• Signal/Alarm LED: A yellow light indicates
either a change in operating status or a
condition that may need response (alarm).
• A flashing light is higher priority than
non-flashing.
Audible signals:
• An audible signal (beep) indicates either a
change in operating status or a condition that
may need response (alarm).
• More frequent beeps indicate higher priority
conditions.
Backlight: A backlight will illuminate the screen
for 15 seconds when the power button is briefly
pressed.
Particle filter: The filters must be always in place
during operation to keep the air going into the
device free of large particles.
Cannula barb: The nasal cannula connects to the
device through this barb.
Power in: Connection for external power from the
AC power supply or DC power cord.
USB port: For service use only.
Page 9 of 272
Cannula
barb
Particle
filter
96-13613-00-01 A
USB
port