WARNING
• The setting of other models or brands of oxygen therapy equipment may not correspond with
the settings of this device.
• The settings of this device may not correspond with the setting for devices that provide
continuous flow oxygen.

PNEUMATIC DIAGRAM

Process flows from left to right

12.3 ELECTROMAGNETIC COMPATIBILITY (EMC) INFORMATION

WARNING
• Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
• Avoid exposure to known sources of EMI (electromagnetic interference) such as diathermy,
lithotripsy, electrocautery, RFID (Radio Frequency Identification), and electromagnetic security
systems such as anti-theft/electronic article surveillance systems, metal detectors. Note that
the presence of RFID devices may not be obvious. If such interference is suspected, reposition
the equipment, if possible, to maximize distances.
• Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the device,
including cables specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
• The device should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the device should be observed to verify normal operation.
If operation is not normal, the device or the other equipment should be moved.
Medical electrical equipment needs to be installed and used according to the EMC information in this
manual.
This equipment has been tested and found to comply with EMC limits specified in IEC 60601-1-2.
These limits are designed to provide a reasonable protection against electromagnetic interference in
a typical home environment.
This concentrator contains Transmitter Module IC: 8595A-NINAB4. Contains FCC ID: XPYNINAB4.
This device complies with Part 15 of the FCC rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause undesired operation.