12.4 GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC
IMMUNITY:
The Concentrator is intended for use in the electromagnetic environment of home, institution, vehicle,
and other transport modalities. The user of the concentrator should make sure it is used in such an
environment. During the immunity testing specified below the Rove 6 will continue to deliver oxygen
within specification.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
EC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. main voltage prior to application of the test level.
©2024 Inogen, Inc. All rights reserved.
IEC 60601 Test Level
3 Vrms
150 kHz to 80 MHz
6Vrms ISM and
amateur frequencies
10V/m
80 MHz
to 2.7 GHz
± 8 kV contact
± 2, 4, 6, 8 and
15 kV air
± 2 kV for power
supply lines
± 1 kV line(s) to
line(s)
0% UT for 0.5 cycle at
0 °, 45 °, 90 °, 135 °,
180 °, 225 °, 270 °, and
315 °.
0% UT for 1 cycle
70% UT for 25/30 cycle
0% UT for 200/300 cycle
30 A/m
Electromagnetic Environment Guidance
The Rove 6 Portable Oxygen Concentrator is
suitable for the electromagnetic environment of
typical home, institution, vehicle, train, airplane,
boat and other transportation environments.
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should
be at least 30%.
Mains power quality should be that of a typical
home, institution, vehicle or other transpiration
and mobile environments.
Mains power quality should be that of a typical
home, institution, vehicle or other transpiration
and mobile environments.
Mains power quality should be that of a typical
home, institution, vehicle and other transportation
and mobile environments. If the user of the Rove 6
requires continued operation during power mains
interruptions, it is recommended that the device
be powered from an uninterrupted power supply.
Power frequency magnetic fields should be at
levels characteristic of a typical home, institution,
vehicle and various mobile environments. Power
frequency magnetic fields from common appli-
ances in the home are not expected to affect
the device.
Page 34 of 272
96-13613-00-01 A