10.1 CANNULA REPLACEMENT
Your nasal cannula should be replaced on a regular basis per the manufacturer's instructions for use.
Consult with your physician and/or equipment provider and/or cannula manufacturer's instructions
for replacement information.
10.2 CASE CLEANING AND DISINFECTION
WARNING
Liquid will damage the internal components of the concentrator and its equipment. To avoid
damage or injury from electrical shock:
• Remove the battery before cleaning
• Turn Off the concentrator and unplug the power cable before cleaning.
• DO NOT allow any cleaning agent to drip inside the air inlet and outlet openings.
• DO NOT spray or apply any cleaning agent directly to the cabinet.
• DO NOT hose down the product.
• DO NOT submerse the device or accessories in liquid
Harsh chemical agents can damage the concentrator and filters.
• DO NOT clean with alcohol and alcohol-based products (isopropyl alcohol), concentrated
chlorine-based products (ethylene chloride), and petroleum-based products or any other
harsh chemical agents.
• Mild liquid dish detergent is recommended.
Periodically clean the case as follows:
1. Make sure the concentrator is off, is removed from the carry bag, and the power cord or battery is
removed.
2. Clean the outside case using a cloth dampened with a mild liquid detergent and water.
3. Allow the concentrator to air dry, or use a dry towel, before returning the concentrator to the carry
bag or backpack and prior to operating the concentrator.
IMPORTANT: The device should receive an external cleaning weekly; accessories should be cleaned
as needed. The device is provided non-sterile and exterior should be cleaned and the output filter
replaced prior to delivering to a new patient.
If desired, disinfection of the exterior surface may be accomplished using an appropriate surface
disinfecting wipe designed to kill bacteria and viruses. If used, follow manufacturers instruction for
use describing application and surface dwell times.
10.3 FILTER CLEANING AND REPLACEMENT (RP-501)
The particle filters must be cleaned weekly to ensure the ease of air flow.
To clean:
1. Remove the battery from the device.
2. Remove the particle filters from both intake ends of the device.
3. Clean the particle filters with a mild liquid detergent and water, rinse in water and dry fully before
reuse.
To purchase additional particle filters, contact your equipment provider or the manufacturer Inogen,
at Inogen.com or by calling +1 877 466 4364, or +31 30 7820689.