Display Icon
7.5.3 LOW PRIORITY ALARM (LEVEL 2)
The following low priority alarms are accompanied by one beep and a solid yellow light.
Display Icon
©2024 Inogen, Inc. All rights reserved.
Description
Remove battery to cool:
Remove battery to cool.
Check battery: Check battery.
Description
Replace columns: Column
replacement is required within
30 days.
Extended start up: Oxygen
concentration is <87% two
minutes after the device's start
up sequence and at least 10
breaths have been detected
within the last minute.
Page 22 of 272
What To Do
The battery needs to be removed
and must be cooled before reuse.
Check the connection of your
battery and ensure that it is properly
attached and latched to the
concentrator. If the battery error
persists with same battery, stop
using the battery and switch to a
new battery or remove the battery
and operate the concentrator using
an external power supply.
What To Do
Contact your equipment
provider to arrange for service
and/or order new columns from
the manufacturer.
Wait a few minutes to see if the
oxygen concentration improves
(alarm will clear). If condition
persists, a secondary alarm will
sound. Follow the instructions for
that alarm or contact your
equipment provider. If alarm
occurs frequently at start up, this
may indicate that maintenance
(column replacement) will soon be
required.